Chemicals and reagents
Ethyl glucuronide (EtG), naltrexone (NTX), 6β-naltrexol (6βNTX), chlordiazepoxide (CDP), norchlordiazepoxide (norCDP), and their deuterated internal standards (ISs), EtG-d5, 6βNTX-d3, CDP-d5, and norCDP-d5, respectively, were purchased from Cerilliant (Austin, TX, USA). The working solutions were prepared by sequentially diluting the standards according to the intended use and storing the diluted solutions at − 20 ℃ until use.
Methanol and water were purchased from J. T. Baker/Avantor (Center Valley, PA, USA), and formic acid was purchased from Sigma-Aldrich (St. Louis, MO, USA). Synthetic urine was purchased from CST Technologies, Inc. (UriSub®, Great Neck, NY, USA).
Urine samples
There is a limitation to validating analytical methods using human urine, because EtG can also be detected in urine following incidental ethanol exposure, including food, medicine, and skin exposure, instead of drinking (Gorgus et al. 2016; Rosano and Lin 2008). Therefore, all validation parameters, except selectivity, were determined using a synthetic urine, ‘UriSub®’. Synthetic urine has very similar physical properties to real urine, such as osmolality, specific gravity, and pH.
Urine samples of 49 probationers who received the medical treatment order requested by the probation office in Busan Metropolitan City and Gyeongsangnam-do Province from 2020 to June 2021 were used to determine alcohol abstinence during alcohol-abuse drug treatment. The samples were stored upon receipt at 4 ℃ and analyzed within 2 weeks.
Instruments
A Cobas C311 immunoassay analyzer (Roche, Hitachi) was used for creatinine analysis using the creatinine Jaffé gen.2 assay (CREJ2, Roche/Hitachi). A URISYS 2400 urine analyzer (Roche, Hitachi) was used to measure the specific gravity of the urine samples. An Agilent 1260 Infinity LC system (Palo Alto, CA, USA) equipped with an AB Sciex QTRAP 4500 MS system (Foster City, CA, USA) was used for LC–MS/MS analysis. A Scherzo SM-C18 (2.0 × 100 mm, 3 µm, Imtakt, Japan) column was used for chromatographic separation. Water containing 0.05% formic acid (mobile phase A) and methanol (mobile phase B) was used as the mobile phases, and the flow rate was 0.2 mL/min. The gradient conditions were as follows: Mobile phase B was maintained at 5% from 0 to 3 min, increased to 40% from 3 to 5 min, maintained at 40% until 7 min, and increased to 95% from 7 to 10 min. Finally, mobile phase B was decreased to 5% from 10 to 10.5 min and maintained at 5% until 20 min to stabilize the LC system. The column temperature was 25 ℃, and the autosampler temperature was 10 ℃.
Electrospray ionization (ESI) was used to generate ions for mass spectrometry. EtG and EtG-d5 were analyzed in the negative-ion mode, whereas the remaining analytes and ISs were analyzed in the positive-ion mode. The ion spray voltage was 5500 V, the ion source temperature was 550 ℃, and ion source gases 1 and 2 were supplied at 55 and 50 (arbitrary unit), respectively. The curtain gas was supplied at 30 (arbitrary unit), and the collision gas was used as the medium. The multiple-reaction monitoring (MRM) mode was used for the quantitative analysis. The retention time (RT), precursor ion, product ions, declustering potential (DP), entrance potential (EP), collision energy (CE), and collision-cell exit potential (CXP) of the standards and ISs were determined.
Sample preparation
Urine samples (100 µL) were mixed with methanol containing IS solutions (300 µL, a mixture of 400 ng/mL for EtG-d5, 40 ng/mL for 6βNTX-d3, 100 ng/mL for CDP-d5, and 200 ng/mL for norCDP-d5) and vortexed for 30 s. Following centrifugation at 30,000 g for 5 min, an aliquot (200 µL) of the supernatant was placed in a test tube for drying. The residue was redissolved in a mixture of methanol ∶ water ∶ formic acid (100 µL, 1∶9∶0.1 (v/v)), and an aliquot (5 µL) of the resulting sample was immediately injected into the LC–MS/MS instrument.
Validation of analytical method
The analytical method was validated for its selectivity, limit of detection (LOD), lower limit of quantification (LLOQ), linearity, accuracy, precision, dilution integrity, recovery (RE), matrix effect (ME), and stability according to the Bioanalytical Method Validation Guidance for Industry (Food and Drug Administration 2018).
Selectivity was confirmed by verifying whether interfering substances influence the retention times of the analytes and the ISs after analyzing 10 different urine samples.
The LOD and LLOQ were measured by determining the signal-to-noise (S/N) ratios using the standard deviations of the signals obtained from 10 urine samples to which a standard material was added, and the noises obtained from 10 blank samples. The LOD was defined as the lowest concentration at which the S/N ratio was 3 or higher, and the LLOQ was defined as the lowest concentration on the calibration curve having an accuracy of ± 20% and a precision of less than 20% while maintaining an S/N ratio of 10 or higher.
The calibration sample concentrations were 5, 15, 50, 100, 250, 500, 800, and 1000 ng/mL for EtG, 6βNTX, CDP, and norCDP, and 1, 3, 10, 20, 50, and 100 ng/mL for NTX. Linearity was confirmed via the correlation coefficient (r). A linear regression model incorporating a weighting factor of 1/x2 was applied to the calibration curves.
The repeatability of the analytical method was confirmed by measuring intra- and inter-day precision and accuracy. For precision and accuracy, quality control (QC) samples of four concentrations (LLOQ, low QC, medium QC, and high QC) were prepared and five samples of each concentration were measured in triplicate. LLOQs were within ± 20% for accuracy and 20% precision. The QCs of three concentrations were within ± 15% for accuracy and 15% precision.
The dilution integrity was evaluated by diluting QC samples corresponding to medium and high concentrations by 5, 10, 20, and 40 times. The diluted samples were divided into five aliquots and analyzed following the pretreatment.
The RE and ME were analyzed after preparing five samples each for sets A, B, and C. Set A was prepared by adding the analyte and the IS to the mobile phase, set B was prepared by adding the analyte and IS to the eluent following the extraction of the blank sample, and set C was prepared by adding the analyte and IS to the blank sample and thereafter extracting it. The recovery (RE = C/B × 100) and matrix effect (ME = B/A × 100) were evaluated by calculating the ratios of the peak areas obtained by analyzing aliquots from each set.
The stability of analytes in urine samples was measured after preparing three samples, each corresponding to low QC and high QC. Bench-top stability was evaluated at room temperature for 24 h, and long-term stability was evaluated at 4 ℃ for 21 days. The autosampler stability was evaluated by re-injecting samples after storing vials containing the samples at 10 ℃ for 4 days.