Table 3 Intra- and inter-day accuracy and precision, recovery (RE), and matrix effect (ME) used to validate the method using quality control (QC) samples

From: Rapid and simple LC–MS/MS determination of urinary ethyl glucuronide, naltrexone, 6β-naltrexol, chlordiazepoxide, and norchlordiazepoxide for monitoring alcohol abuse

Compound

QC sample (ng/mL)

Intra-day (n = 5)

Inter-day (n = 15)

RE (n = 6)

%

ME (n = 6)

%

% Bias

% CV

% Bias

% CV

EtG

5

 − 4.7

3.8

 − 8.4

8.1

89.7

99.6

15

 − 1.4

6.7

 − 0.2

7.1

90.6

102.0

250

 − 1.3

1.8

2.0

6.3

97.7

96.0

800

 − 0.3

0.9

3.4

6.8

98.0

98.3

NTX

1

7.5

6.5

7.8

6.8

95.9

52.2

3

 − 2.3

6.7

 − 5.3

5.2

86.9

49.0

20

 − 2.2

6.3

 − 3.8

6.8

103.0

50.3

80

11.4

1.1

8.3

5.3

93.1

52.0

6βNTX

5

 − 7.8

9.5

 − 2.7

15.2

95.0

89.2

15

 − 7.3

7.2

 − 1.0

10.1

88.6

88.1

250

 − 4.3

5.3

 − 0.7

7.6

103.9

77.6

800

0.1

4.3

 − 4.4

6.7

104.2

71.7

CDP

5

11.2

5.2

2.3

10.0

89.7

48.9

15

5.9

6.6

2.7

7.8

93.8

46.3

250

 − 0.4

7.0

1.3

5.9

96.6

47.0

800

3.9

6.6

2.2

7.2

91.2

49.4

norCDP

5

 − 5.0

5.3

 − 0.5

6.3

95.3

45.3

15

1.3

6.1

 − 0.3

4.6

90.1

42.9

250

8.8

3.0

2.8

5.6

103.0

42.1

800

1.0

2.1

0.8

3.7

100.6

43.0