Method validation and measurement uncertainty of possible thirty volatile organic compounds (VOCs) presented in the polyethylene present in bottled drinking waters sold in Turkey

Abstract

This study was actualized for the simultaneous determination of possible thirty VOCs presented in drinking waters in Turkey by direct injection to purge and trap (PT) gas chromatography-mass spectrometry (GC-MS). It consists of selectivity, linearity, the limit of detection (LOD) and limit of quantification (LOQ), accuracy (recovery), precision, trueness, and measurement uncertainty studies. In linearity, the values of correlation coefficients (r2) for the matrix-matched calibration curves were higher than 0.998 for all analytes. This method showed high sensitivity (LOD: 0.011–0.040 μg/L; LOQ: 0.035–0.133 μg/L), quite sufficient recovery (82.6% to 103.1%) for accuracy, and acceptable precision (intra-day recovery: 81.5–104.4%, relative standard deviation (RSD): 1.04–9.81%; inter-day recovery: 92.6–104.1%, RSD: 1.15–7.52%). All the recovery and RSD values obtained below 10% are evaluated agreeable in point of the AOAC and EURACHEM/CITAC validation guidelines. The recovery percentages of all analytes in CRM changed between 80.3 and 109.9% and the RSD (%) values for each analyte obtained below 10%. The proficiency test results were satisfactory and comparable (z score less than or equal to 2.0 is no questionable or satisfactory) to those obtained by other laboratories participating in the round. The calculated percentage of relative uncertainties for each analyte changed from 2.99 to 10.10% and the major contribution to uncertainty budget arises from the calibration curve and repeatability. Therefore, the results demonstrate that this method is applicable for the determination of possible thirty VOCs in drinking waters in routine analysis for custom laboratories.

Introduction

Global warming caused by the increase in greenhouse gas emissions and increasing urbanization and industrialization activities has adverse effects on water and water resources such as drought. Nowadays, this situation is reaching much serious level in terms of threating the quality of human life. Therefore, protection from all kinds of pollution and more efficient use of existing water resources are of vital importance for human health (Dehghani Darmian et al. 2018). The presence of chemical substances in the water environment causes many diseases and health problems such as cancer, liver, and kidney diseases (Tehrani and Van Aken 2014). Various regulations have been published at national and international levels regarding the protection of water such as European Union (EU) Water Framework Directive (2000/60/EC), Drinking Water Directive (98/83/EC), Prevention and Control Regulation (PCR), the Food and Environmental Protection Act (FEPA/1985), the Control of Pesticides Regulations (COPR/1986), and the United Kingdom (UK) Surface Waters (dangerous substances) Regulations (SI 1997/2560). These regulations ensure that water is continuously monitored and kept under control (Tombs 2000).

Volatile organic compounds (VOCs) have been one of the most studied organic pollutant class in recent years due to its persistent and toxic effects on human health. For example, methylbenzene (toluene), dichlorobenzenes, trichlorobenzenes, dimethyl benzenes (xylenes), tetrachloromethane, dibromochloromethane, and trichloroethene are persistent organic substances for the environment and human life (Tombs 2000; Jurdakova et al. 2008; Bhattacharya et al. 2016; Mirzaei et al. 2016). They are composed of carbon chains and possess high vapor pressure at room temperature (Kountouriotis et al. 2014). The American Society for Testing and Materials (ASTM) has stated in pursuant of D3960 test method that VOCs have higher vapor pressure than 0.1 mmHg (Table 1) at 25 °C (ASTM 1989). Their boiling points change between 40 and 260 °C (Güzel et al. 2018). Besides, World Health Organization (WHO) describes compounds with boiling point of 50 °C to 100 °C as very volatile organic compounds (VVOC) (WHO 1987). High vapor pressure (low boiling point) of VOCs allows greater number of molecules to move freely and causes the substance to change form quickly. This allows the substances to spread easily into the environment. For this reason, they can make carcinogenic, mutagenic, and toxic effects by easily reaching people (Tsuchiya 2010). Hence, they need to analyze their species and amounts, especially in water in order to reduce the toxic effects of human life.

Table 1 Some physical, chemical, and GC-MS properties of VOCs (ATSDR 1997; EPA 2002; Alonso et al. 2011; Güzel et al. 2018)

Analytical chemistry and food laboratories need to sensitive and reliable methods more than ever to obtain practicable qualitative and quantitative data. Recently, chemical analysis has been taken into consideration as the main approach for the detection of VOCs and similar organic pollutants in water samples. Because of their toxicity and persistence, today, an analytical method is required to determine trace amounts in the water matrix. For this purpose, this study was to validate a fully automated analytical method for the determination of thirty water-soluble VOCs in drinking water samples using purge and trap (PT) gas chromatography-mass spectrometry (GC-MS) system. Conventional measurement methods include pretreatment stage to concentrate the sample before analysis. These pre-treatments require a long period of time and are costly due to the use of solvents. In contrast to these methods, the proposed method is completely automatic and does not include an additional pre-treatment step and solvent use. The sample has been concentrated by PT system. This study especially provides saving on time and reduces the cost of analysis. The method, that has been optimized and has been identified measurement uncertainty limits in detail, contributes rapid, simple, sensitive, and accurate qualification and quantification of possible thirty VOCs in drinking water samples at the same time.

Materials and methods

The studies of this research were carried out in the Scientific and Technological Research Council of TURKEY Marmara Research Center (TUBITAK MAM) Environment and Cleaner Production Institute laboratories. They possess national accreditation certificate taken from Turkish Accreditation Agency (TURKAK) pursuant to TS EN ISO/IEC 17025:2012 standard since July 16, 2010, and “Environmental Measurement and Analysis Qualification Certificate” from the Republic of Turkey Ministry of Environment and Urbanization acquired on February 21, 2011, respectively. Besides, they have international accreditation certificates acquired from German Accreditation Council DAR/DAP (Deutscher Akkreditierung Rat) between December 17, 2002 and 2010.

Reagents and chemicals

Ampoule of VOC standard solution containing 60 different chemicals (200 mg/L each in methanol) was purchased from High-Purity Standards Co., Inc. (North Charleston, USA). Standard solution of 1,2,3-trichlorobenzene in methanol (GC gradient grade) with purities higher than 99.8% was supplied by Dr. Ehrenstrofer (Augsburg, Germany). 4-Bromofluorobenzene (25 mg/mL in methanol), which is used as the internal standard (IS) in the analysis, was purchased from Absolute Standards Co., Inc. (Hamden, USA). The highest purity grade methanol was obtained from Merck (Darmstadt, Germany) and the purification of water was performed with Milli-Q Plus system (EMD Millipore, Billerica, MA).

Further, 1 mg/L and 10 mg/L concentrations of stock standard solutions were prepared by mixing 200 mg/L stock VOC solution and 100 mg/L 1,2,3-trichlorobenzene. These solutions were stored in a freezer (− 20 °C) at 1.5 mL vials. They should be prepared again once a month because of their stability. Standard solutions were made ready for the linearity and other studies by dilution of stock standard solutions in the ultra-purified water.

Water samples

Commercial one hundred drinking water samples sold in grocery markets and supermarkets in Turkey were supplied used in the study. They were preserved and handled in accordance with the related International Organization for Standardization (ISO) standard (ISO 2012). The analysis of whole samples (stored at less than 5 °C) was carried out within 48 h after they were purchased.

Analytical instrumentation

PT analytical condition

PT system used for the extraction and enrichment of investigated VOCs from samples was an OI Analytical Eclipse model 4660 sample concentrator (College Station, TX, USA) equipped with an OI Analytical 4552 water/soil autosampler. The sample was added into 50 mL PT glass vial until filled and then, 25 mL of sample was transferred to the purge unit, which is the extraction cell of the analytes in the sample. The IS (20 μL from IS with a concentration of 2.5 mg/L) was added to the sample via injector before purge process. To extract all the volatile analytes wholly, a sample was purged with the help of pure helium gas (purge gas). At the end of the purging process, analytes were sent to the trap unit (trapping material is Purge Trap K VOCARB 3000). Consequently, collected analytes in the trap were transferred to GC column for the separation by desorbing via high temperature. The PT system was programmed as follows: purge-ready temperature: 35 °C, purge time: 11 min, dry purge time: 5 min, purge gas flow rate: 50 mL/min, desorb preheat: 180 °C, desorb time: 1 min, desorb temperature: 180 °C, bake time: 7 min, bake temperature: 260 °C.

GC-MS analytical condition

The analyses for the determination and quantification of thirty VOCs were actualized by using Agilent Technologies 6890N Network GC system (Avondale, USA) coupled with a 5975C inert mass spectrometer with a triple detector (MSD). The separations were performed using a DB-5ms fused silica capillary column as an analytical column in GC separation having 60 m × 0.25 μm with a 0.25 μm film thickness (Agilent Technologies J&W Scientific, Folsom, CA, USA) and the injector temperature was set to 200 °C. The split mode for all analysis was used with a ratio of 20:1 by using a 4.0 mm split liner at 200 °C. The helium of 99.9995% purity at a gas flow rate of 1.0 mL/min was employed as the carrier gas with a pressure of 110 kPa in the port of injection. The adjustment of initial temperature and ionization voltage of GC system was adjusted to 280 °C and 70 eV by selecting electron impact ionization (EI-SIM). At the beginning of the analysis, the column temperature of GC system was programmed to 40 °C for 2 min and was run to increase to 200 °C at a rate of 7 °C/min. When the GC system arrived at the final temperature, it was held for 5 min. The MS was arranged to a total-ion-monitoring mode (m/z: 25–550). The confirmation of each compound was realized with two MS characteristic ions; the ratio of two MS characteristic ions and GC retention times matches to the known standard compounds. The confirmation and quantification of each VOC were shown in Table 1 with relevant target and quantified ions. The optimum performance values of GC-MS and PT systems mentioned above for the analysis of related VOCs in water were determined in consequence of experimental works carried out before the validation study of these parameters.

Results and discussion

Method optimization study

Optimization of PT condition

PT is a sample pre-treatment system that allows the collection of VOCs in the sample simply with high precision and efficiency. In PT system, VOCs found in water are transported by boiling efficiently to the vapor phase and they are passed through an absorbent trap with the aid of the purge gas such as helium. After purging process has finished, the trap containing VOCs is heated to deliver a gas chromatographic column and VOCs along with heating are delivered with the carrier gas. At this stage, the parameters of purge time, purge gas flow rate, and desorb temperature play vital role for the most effective collection of VOCs from the sample.

In the determination of VOCs, purge gas flow rate and purge have a crucial importance for the efficiency of the trap and analysis time. The high flow rate brings about both failure to capture VOCs by the trap and low MS sensitivity and troubled drying because of blowing-out of water. Besides, the low flow rate affects the collection of VOCs from the sample per unit time and leads to prolonged analysis time (Hong-Hai et al. 2015). Thus, the optimum value of the flow rate of purge gas must be determined to actualize the accurate and sensitive analysis. Measurements at different flow rates with parallel study (n = 4) were performed for 5 μg/L standard solution at a certain purge time. Figure 1a presents the relationship between flow rate of purge gas and related peak area. The results demonstrated that 40 mL/min of purge gas flow with regard to the high purge efficiency was found optimum for almost all of VOCs with negligible exceptions such as trichloromethane, tetrachloromethane, isopropylbenzene, 1,2,4-trichlorobenzene, and naphthalene. After establishing the optimum value of the purge gas flow rate, the purge time was determined. Measurements with different purge times (8, 9, 10, 11, 12, and 13 min) were carried out for 5 μg/L standard solution at the purge gas flow rate of 40 mL/min. Fig. 1b shows the dealings between purge time of instrument and peak area of VOCs. When the peak areas of VOCs in the sample were examined, the results revealed that purge time of 11 min is more than enough to get the best results. Exposing the sample to a longer purge time will not cause any appreciable change in results. It prolongs the analysis time and leads to unnecessary gas consumption. Thus, 11 min was decided as fittest purge time in real work.

Fig. 1
figure 1

Effect of a) purge gas flow rate, b) purge time, and c) desorb temperature on analytical measurements

At the end of purging of VOCs in sample, VOCs retained in trap, which allows the purge gas to pass through to vent, were then desorbed by heating the trap at a suitable desorb temperature. Because, overheating the trap may cause decomposition of the compounds and adsorbent material. Therefore, the optimization of desorb temperature was actualized in this work. Measurements went through with 5 μg/L standard solution at the purge time of 11 min and at the purge flow rate of 40 mL/min. The results in Fig. 1c exhibited that almost whole of VOCs in sample was collected from the trap at 180 °C of desorb temperature, which was found optimum.

Method verification study

The performance of proposed method was evaluated with regard to selectivity, linearity, the limit of detection (LOD) and quantification (LOQ), trueness, accuracy, and precision in accordance with Commission Decision EURACHEM Guideline (EURACHEM/CITAC 2014) and Guidelines for Standard Method Performance Requirements (AOAC 2016). Water having high purity was used as a blank sample.

Selectivity

The selectivity study of the method was carried out by the analyses of seven drinking water samples to control the stability of retention times and the ratio of target and quantified ion signals of each analytes. In consequence of GC-MS analyses of these samples, the results indicated that there are no peaks of interfering compounds that will cause positive results in the analysis.

Linearity

For the evaluation of linearity of the calibration curve, matrix-match calibration curves of spiked drinking water samples were employed. The calibration curves for each VOCs were prepared at nine (0.15 μg/L, 0.25 μg/L, 0.5 μg/L, 1 μg/L, 2 μg/L, 5 μg/L, 10 μg/L, 20 μg/L, and 40 μg/L) concentration levels (Fig. 2) spiked from two stock standard solutions(1 mg/L and 10 mg/L) by paying regard to their signal intensities. The values of correlation coefficient (r2), calibration curve equation (y = ax + b) and dynamic linear range from all calibration curves are illustrated in Table 2. In a similar study, Ferreira et al. (2016) presented that the linear regression with correlation coefficient should be equal or better than 0.99 (good linearity) for the evaluation of the linearity range. In the light of this approach, the values of correlation coefficients for the matrix-matched calibration curves were higher than 0.998 for all analytes in this study. The analytical response linearity in the working concentration range can be assessed as a great in pursuance of correlation coefficients. The dynamic linear range was determined from 0.15 to 20 μg/L for the majority of the studied VOCs except for dichloromethane, toluene, and p&m-xylene.

Fig. 2
figure 2

Overlaid chromatograms at nine concentration levels for each VOCs

Table 2 Linearity, LODs, LOQs, and accuracy study of VOCs

LOD and LOQ

The evaluation of method sensitivity has been done with the determination of LOD and LOQ values. Figure 3 shows the chromatogram of 5 μg/L standard solution, which is used for the determination LOD and LOQ levels of VOCs. The LOD and LOQ were calculated as (Eq. (1) and Eq. (2)):

Fig. 3
figure 3

Chromatogram of 5 μg/L standard solution

$$ LOD=\frac{3\upsigma}{\mathrm{S}} $$
(1)

and,

$$ LOD=\frac{10\upsigma}{\mathrm{S}} $$
(2)

where σ is the standard deviation and S is the slope of the standard curve.

The calculations of LOD and LOQ results were made from the mean noise value in the chromatogram. The results of LOD and LOQ are given in Table 3. Thus, LOD values for all VOCs were found to be in the range of 0.011 μg/L and 0.040 μg/L, and the ranges of LOQ for VOCs were obtained from 0.035 to 0.133 μg/L. The LOD and LOQ values presented in this work are similar to LODs and LOQs reported for VOCs analysis in drinking water samples and similar types of water samples developed for Jurdakova et al. (2008), Chary and Fernandez-Alba (2012), and Ueta et al. (2013). Moreover, The LOD and LOQ values of some VOCs studied in this work are less than the values obtained in the other studies (Kubinec et al. 2004; Alonso et al. 2011; Chen et al. 2015) in literature for the analysis of VOCs in water samples.

Table 3 Intra and inter-day precision results of VOCs

The accuracy of the method

The reliability of the method was confirmed by recovery experiments performed in the drinking water sample spiked with three different concentrations of 1 μg/L, 2 μg/L, and 5 μg/L for VOCs and seven measurements were carried out for each spiked concentration level. The recovery of each analyte at each concentration level was calculated by using Eq. (3):

$$ \mathrm{Analyte}\ \mathrm{recovery}\left(\%\right)=\frac{\mathrm{Determined}\ \mathrm{concentration}\ \mathrm{of}\ \mathrm{analyte}}{\mathrm{Spiked}\ \mathrm{concentration}\ \mathrm{of}\ \mathrm{analyte}}x100 $$
(3)

The percentage of mean recovery and the relevant RSDs for each VOC recovery at each spiked concentration level were depicted in Table 2. The recoveries ranged from 82.6 to 103.1%. The percentage of RSD values for VOCs changed between 2.48 and 9.19. These results demonstrate that this method has considerably good and sufficient capability for the accurate analysis of VOCs in water. As can be seen, the accuracy results of the method were satisfied and verified with validation guideline for Standard Method Performance Requirements (AOAC 2016). Besides, Wu and Fung (2010) for BTEX (benzene, toluene, ethylbenzene, xylenes) (94.7–99.1%) and Ueta et al. (2013) for BTEX and THMs (trihalomethanes) (97.6–103.9%) reported similar recoveries.

The precision of the method

The precision of the method was calculated as RSD and was assessed as intra-day and inter-day precision. Within the scope of intra-day precision (repeatability), six drinking water samples were spiked with VOCs at each fortification level (2 μg/L and 5 μg/L) on the same day. The inter-day precision (reproducibility) study was done by injecting each of 5 μg/L concentrations of standard solution six times in different days in drinking water sample. The chromatograms of precision study were shown in Fig. 4. As can be seen in Table 3, RSD values for 2 μg/L and 5 μg/L concentration levels in intra-day precision were in the range of 1.04% to 9.81% and 1.67% to 7.04%, respectively. RSD values for 5 μg/L concentration levels in inter-day precision changed between 1.40 and 7.52%. All the RSD values obtained below 10% are considered acceptable for this parameter in accordance with the validation guideline (AOAC 2016).

Fig. 4
figure 4

Overlaid chromatograms of precision study at 2 μg/L and 5 μg/L standard solutions

Trueness

The performance of a method is measured by evaluating with regard to the precision and trueness. From this perspective, trueness refers to how close the average result (obtained from the method) of a series of studies is to the actual value (EURACHEM/CITAC 2014). The trueness of analytes in this study was checked with PriorityPollutnT QC 710 coded certified reference material (CRM) issued from ERA-A Waters Company. The results of relevant CRM measured by PT-GC/MS system are presented in Table 4. The recovery percentages of all analytes changed between 80.3 and 109.9% and the RSD (%) values for each analyte obtained below 10%. These results are favorable according to the relevant verification guidelines (EURACHEM/CITAC 2012; EURACHEM/CITAC 2015).

Table 4 CRM and inter-laboratory test results for VOCs

The proficiency test is an important study to evaluate the validity of verified analysis methods of all the laboratories. It is also performed to ensure an objective assessment of the accuracy and reliability of laboratory calibration, test, and analysis results. Therefore, the proposed method was applied in the proficiency test for each VOCs studied in water matrix to evaluate the validation of the method. The results obtained from the proficiency test are given in Table 4. It was evaluated laboratory performance based upon the z score, which was calculated as (Eq. (4)):

$$ z=\frac{\left(x-X\right)}{\mathrm{SDPA}} $$
(4)

where x is the laboratory reported value and X is the assigned value, determined by formulation or robust mean and SDPA is the standard deviation for proficiency assessment on a fixed percentage.

The obtained z scores in the test may be categorized in accordance with the literature (Camino-Sánchez et al. 2013) as follows: (a) z score less than or equal to 2.0 is no questionable or satisfactory; (b) z score greater than 2.0 and less than 3.0 is warning; and (c) z score greater than or equal to 3.0 is unsatisfactory. As a result, the validity of the method was proved by the fact that the results obtained in Table 4 were satisfactory and comparable to those obtained by other laboratories participating in the round. The results indicate that the method is not affected by a significant systematic error that may cause incorrect results, except for random error. In conclusion, both results of CRM and proficiency test indicate that this method is rather good and sufficient capability for the rapid and sensitive determination of possible thirty VOCs in drinking waters.

Measurement uncertainty

The study about method validation has an important role in explaining the reliability and accuracy of method results. However, this is not enough by oneself to commentate and compare the results accommodately. It should be stated the measurement uncertainty as well as the method validation (Rozet et al. 2011). The uncertainty of VOCs in drinking water samples was decided by using PT-GC/MS. The uncertainties in the study were used as parameters to identify the standard combined uncertainty, expanded uncertainty, and relative uncertainty based on the related guidelines (EURACHEM/CITAC 2012; EURACHEM/CITAC 2015). Therefore, the following individual source parameters were taken into account for the calculation of the uncertainties of VOCs in this study: (a) standard preparation (Std), (b) sample preparation (Sample), (c) calibration curve (Cal), and (d) repeatability (Rep).

  • In compliance with these individual uncertainty sources, the standard combined uncertainty of the VOCs is calculated with the following equation (Eq. (5)):

$$ \frac{u_{\mathrm{combined}}\ }{Conc}=\sqrt{{\left(\frac{\mathrm{u}(Std)}{Std}\right)}^2+{\left(\frac{\mathrm{u}\left(\mathrm{Sample}\right)}{\mathrm{Sample}}\right)}^2+{\left(\frac{\mathrm{u}(Cal)}{Cal}\right)}^2+{\left(\frac{\mathrm{u}(Rep)}{Rep}\right)}^2} $$
(5)
  • Calculation of the uncertainty derived from calibration curve, u(Cal) (Eq. (6)):

$$ u\left({c}_o\right)=\frac{S}{B_1}+\sqrt{\frac{1}{p}+\frac{1}{n}+\frac{{\left({c}_o-{c}_{\mathrm{average}}\right)}^2}{S_{xx}}\ }\ {S}_{xx}=\sum \limits_{i=1}^n{\left({c}_i-{c}_{\mathrm{average}}\right)}^2 $$
(6)

where S is the standard deviation, B1 is the slope, p is the number of measurements to determine co, n is the number of measurements for the calibration, co is the analyte concentration in drinking water sample, caverage is the mean value of the different calibration standards (n number of measurements), i is the index for the number of measurements to obtain the calibration curve, and ci is the individual calibration standard value obtained from the calibration equation.

  • Calculation of the uncertainty derived from repeatability, u(Rep):

The repeatability study was explained by the standard deviation of the results of eight consecutive measurements for each VOCs on the same day. Uncertainty from reproducibility was estimated in accordance with Aslan-Sungur et al. (2014) using the following equation (Eq. (7)):

$$ u(Rep)=\frac{RSD}{\sqrt{n}} $$
(7)

where RSD is the relative standard deviation and n is the number of measurement (in this case, n = 8).

  • Calculation of the uncertainty derived from standard preparation, u(Std):

Standard preparation from stock VOC solution (200 mg/L) for the calibration curve is one of fundamental source to the uncertainty budget. The preparation of standards was done by using pipette (pip) (100 μL and 1000 μL) and volumetric flask (flask) (50 mL). Thus, the uncertainty calculation of standard preparation (Eq. (8)) includes the sum of the uncertainties of stock solution and glassware used.

$$ \frac{u_{(Std)}\ }{C_{(Std)}}=\sqrt{{\left(\frac{u(pip)}{V_{(pip)}}\right)}^2+{\left(\frac{u\left(\mathrm{stock}\right)}{V_{\left(\mathrm{stock}\right)}}\right)}^2+{\left(\frac{u\left(\mathrm{flask}\right)}{V_{\left(\mathrm{flask}\right)}}\right)}^2} $$
(8)

where u(pip) is the pipette uncertainty and u(stock) is the uncertainty of the stock VOC solution. u(pip) (Eq. (9)) is calculated from the square root of the sum of the squares of both the uncertainty of calibration of the pipette [u(pipcal)] and the uncertainty of temperature effect [u(temp)]. The calculation of u(flask) (Eq. (10)), which is defined as the preparation of the standard solution by using a flask, is carried out the combination of the uncertainty of calibration of the flask [u(flaskcal)] and the uncertainty of temperature effect [u(temp)].

$$ {u}_{(pip)}=\sqrt{{\left(\mathrm{u}\left(\mathrm{pipcal}\right)\right)}^2+{\left(\mathrm{u}\left(\mathrm{temp}\right)\right)}^2} $$
(9)
$$ {u}_{\left(\mathrm{flask}\right)}=\sqrt{{\left(\mathrm{u}\left(\mathrm{flask}\mathrm{cal}\right)\right)}^2+{\left(\mathrm{u}\left(\mathrm{temp}\right)\right)}^2} $$
(10)

Temperature effect means that the temperature variation in the laboratory was commonly accepted as ± 3 °C in (EURACHEM/CITAC 2012).

$$ u\left(\mathrm{temp}\right)=\frac{3\ x\ V\ x\ Q}{1.73} $$
(11)

where u(temp) is the standard uncertainty of temperature effect, V is the measured volume of equipment such as 1000 μL for pipette, and Q is the volume expansion coefficient of the methanol (Qmethanol: 1.49 × 10−3 °C−1).

  • Calculation of the uncertainty derived from sample preparation, u(Sample):

Fifty-milliliter flask was used to take the sample volume and this method includes solventless sample preparation technical procedure. For this reason, the flask uncertainty (u(flask)) in (Eq. (10)) is the source of u(Sample).

Table 5 shows the overview of the contributions of the individual uncertainty sources to the measurement of investigated VOCs at 5 μg/L. The calculation of the combined and expanded uncertainties (μg/L) (based on a 95% confidence level using a coverage factor (k) of 2) of VOCs was performed for nine standards from 0.15 to 40 μg/L used in calibration. The calculated percentage of relative uncertainties for each analyte changed from 2.99 to 10.10% and it is clear that among the four sources of uncertainty, the major contributions to the uncertainty budget consist of the calibration curve and repeatability. The contribution (%) of the standard preparation to the to the measurement uncertainty budget changed from 0.57 and 1.69%, and the contribution (%) of the sample preparation to the measurement uncertainty budget ranged from 3.38 and 10.03%. Their results indicate the contributions of standard preparation and sample preparation on the uncertainty budget are almost negligible to the combined standard uncertainties.

Table 5 Measurement uncertainty study of investigated VOCs at 5 μg/L

Application to real samples

The proposed analytical method in this study was used to monitor the level of VOC residues in sixteen different drinking water brands obtained from the supermarket. More than one hundred sixty drinking water samples were analyzed with the related analytical method. In Turkey, the presence of VOCs in drinking water has been determined mostly on the basis of the thresholds set out in the European Union's Drinking Water Directive (98/83/EC), the relevant international (e.g., WHO) or national standards. Table 6 shows the limit values of relevant national and international regulations for VOCs and their observed concentration ranges in the drinking water samples. Prior to analysis of real samples, the control of retention times, molecular and fragment ions of all analytes were performed by measuring a known standard of concentration. Furthermore, the retention times, target, and fragment ions of the compounds detected as positive in the chromatograms of these samples were checked by comparing the chromatographic peaks of the relative standards. The examination of total ion chromatograms of standard and actual samples was also done in detail. There were no interfering compounds that would cause false positive results in the chromatogram. In this study, dichloromethane, trichloromethane, 1,2-dichloroethane, benzene, toluene, tetrachloroethene, tetrachloromethane, ethylbenzene, tribromomethane, 1,1,2,2-tetrachloroethane, naphthalene, and hexachloro-1,3-butadiene were detected at low concentrations in the polyethylene bottled waters. The concentrations of these compounds had reported similar results in the literature (Leivadara et al. 2008; Al-Mudhaf et al. 2009; Ikem 2010). These substances were positively detected in forty-eight drinking water samples within the ranges summarized in Table 6 and there were no individual VOC parameters exceeding national and international limits (EU 1998; RCWHC 2005; WHO 1993; WHO 2001). Six-point calibration curves by fitting its area ratio were applied for the quantification of analytes.

Table 6 Limit values of national and international regulations for VOCs and their observed concentration ranges in the drinking water samples (EU 1998; RCWHC 2005; WHO 1993; WHO 2001)

Conclusion

The verification study of proposed method has been performed for the simultaneous analysis of the presence of possible thirty VOCs in drinking waters by direct injection to PT-GC/MS in compliance with national and international legislations such as European Union’s Drinking Water Directive (98/83/EC) regarding the quality of polyethylene bottled waters determined to human consumption. This method includes solventless sample preparation technical procedure, which is called PT system. It has been validated in pursuant to ISO/IEC 17025:2017 requirements, following the AOAC and EURACHEM/CITAC validation guidelines with the assessment of selectivity, linearity, LOD and LOQ, accuracy (recovery), precision, trueness, and measurement uncertainty studies. In linearity of calibration curves, the values of correlation coefficients for the matrix-matched calibration curves were higher than 0.998 for all analytes in this study. The analytical response linearity in the working concentration range can be assessed as a great in pursuance of correlation coefficients. The dynamic linear range was determined from 0.15 to 20 μg/L for the majority of the studied VOCs except for dichloromethane, toluene, and p&m-xylene. The LOQ values were found to be in the range of 0.011 μg/L and 0.040 μg/L and the ranges of LOQ for VOCs were obtained from 0.035 to 0.133 μg/L, respectively. These results indicate high sensitivity of the method. This method reveals quite sufficient recovery (82.6% to 103.1%) for accuracy; acceptable precision (intra-day recovery: 81.5–104.4%, RSD: 1.04–9.81%; inter-day recovery: 92.6–104.1%, RSD: 1.15–7.52%). All the recovery and RSD values obtained below 10% can be interpreted agreeable with respect to related validation guidelines. The trueness of method was evaluated with CRM and through participation in proficiency testing analyses. At the end of analysis of CRM, the recovery percentages of all analytes varied from 80.3 and 109.9% and the RSD (%) values for each analyte obtained below 10%. The results are acceptable in comparison to certified value. Moreover, there were no obtained questionable or unsatisfactory z score results in the proficiency test. In the measurement uncertainty, the calculated percentage of relative uncertainties for each analyte was changed from 2.99 to 10.10% and the calibration curve and repeatability possess powerful impacts on the combined uncertainty. On account of the high sensitivity and good accuracy obtained with the combination of PT and single quadrupole MS, developed and validated screening method were effectively proven to work and allows the analysis of thirty VOCs at low concentrations by applying drinking water samples.

Availability of data and materials

Research data have been provided in the manuscript.

Abbreviations

ASTM:

American Society for Testing and Materials

COPR:

Control of Pesticides Regulation

CRM:

Certified Reference Material

DAR:

Deutscher Akkreditierung Rat

EU:

European Union

FEPA:

Food and Environmental Protection Act

GC-MS:

Gas Chromatography-Mass Spectrometry

IS:

Internal Standard

ISO:

International Organization for Standardization

LOD:

Limit of Detection

LOQ:

Limit of Quantification

PCR:

Prevention and Control Regulation

PT:

Purge & Trap

RSD:

Relative Standard Deviation

TUBITAK MAM:

The Scientific and Technological Research Council of TURKEY Marmara Research Center

TURKAK:

Turkish Accreditation Agency

UK:

United Kingdom

VOCs:

Volatile Organic Compounds

VVOC:

Very Volatile Organic Compounds

WHO:

World Health Organization

References

  • Al-Mudhaf HF, Alsharifi FA, Abu-Shady ASI. A survey of organic contaminants in household and bottled drinking waters in Kuwait. Sci Total Environ. 2009;407(5):1658–68.

    Article  CAS  Google Scholar 

  • Alonso M, Cerdan L, Godayol A, Antico E, Sanchez JM. Headspace needle-trap analysis of priority volatile organic compounds from aqueous samples: application to the analysis of natural and wastewaters. J Chrom A. 2011;1218:8131–9.

    Article  CAS  Google Scholar 

  • AOAC. (2016) Guidelines for standard method performance requirements. Appendix F. Association of Analytical Chemists. 20th ed. vol. 2.

    Google Scholar 

  • Aslan-Sungur G, Gaga EO, Yenisoy-Karakas S. Estimation of measurement uncertainty of polychlorinated biphenyls, polycyclic aromatic hydrocarbons and organochlorine pesticides in the atmosphere using gas chromatography-mass spectrometry and gas chromatography-electron capture detector. J Chrom A. 2014;1325:40–8.

    Article  CAS  Google Scholar 

  • ASTM. The ASTM standard practice for determining volatile organic compounds (VOC) contents of paints and related coating (D3960). Philadelphia: American Society for Testing and Materials; 1989.

    Google Scholar 

  • ATSDR. Agency for toxic substances and registry. Atlanta: U.S. Public Health Service. U.S. Department of Health and Human Service; 1997.

    Google Scholar 

  • Bhattacharya SS, Kim KH, Ullah MA, Goswami L, Sahariah B, Bhattacharyya P, Cho SB, Hwang OH. The effects of composting approaches on the emissions of anthropogenic volatile organic compounds: a comparison between vermicomposting and general aerobic composting. Environ Pollut. 2016;208:600–7.

    Article  CAS  Google Scholar 

  • Camino-Sánchez FJ, Zafra-Gómez A, Ruiz-García J, Vílchez JL. Screening and quantification of 65 organic pollutants in drinking water by stir bar sorptive extraction-gas chromatography-triple quadrupole mass spectrometry. Food Anal Methods. 2013;6:854–67.

    Article  Google Scholar 

  • Chary NS, Fernandez-Alba AR. Determination of volatile organic compounds in drinking and environmental waters. Trends Anal Chem. 2012;32:60–75.

    Article  CAS  Google Scholar 

  • Chen PS, Tseng YH, Chuang YL, Chen JH. Determination of volatile organic compounds in water using headspace knotted hollow fiber microextraction. J Chrom A. 2015;1395:41–7.

    Article  CAS  Google Scholar 

  • Dehghani Darmian M, Monfared SAH, Azizyan G, Snyder SA, Giesy JP. Assessment of tools for the protection of the quality of water: uncontrollable discharges of pollutants. Ecotoxicol Environ Saf. 2018;161:190–7.

    Article  CAS  Google Scholar 

  • EPA. Volatile organic compounds in water, soil, soil gas, and air by direct sampling ion trap mass spectrometry (DSITMS). USA: U.S. Public Health Service. U.S. Department of Health and Human Service; 2002.

    Google Scholar 

  • EU. (1998) The quality of water intended for human consumption. Drinking Water Directive. Council Directive 98/83/EC.

    Google Scholar 

  • EURACHEM/CITAC. In: LGC, editor. Quantifying uncertainty in analytical measurement. 3rd ed. Teddington; 2012.

  • EURACHEM/CITAC. The fitness for purpose of analytical methods: a laboratory guide to method validation and related topics. 2nd ed. Teddington: LGC; 2014.

    Google Scholar 

  • EURACHEM/CITAC. Setting and using target uncertainty in chemical measurement. 1st ed. Teddington: LGC; 2015.

    Google Scholar 

  • Ferreira JA, Ferreira JMS, Talamini V, Facco JF, Rizzetti TM, Prestes OD, Adaime MB, Zanella R, Bottoli CBG. Determination of pesticides in coconut (Cocos nucifera Linn.) water and pulp using modified QuEChERS and LC-MS/MS. Food Chem. 2016;213:616–24.

    Article  CAS  Google Scholar 

  • Güzel B, Canli O, Öktem OE. Volatile organic compounds in waters and their health effects. Anadolu Uni J Sci and Tech C - Life Sci and Biotech. 2018;7(2):277–90.

    Google Scholar 

  • Hong-Hai Z, Jian-Long L, Gui-Peng Y, Yu-Chen S, Na J. Purge- trap gas chromatography and mass spectrometric method for analysis of isoprene in natural waters. Chin J Anal Chem. 2015;43(3):333–7.

    Article  Google Scholar 

  • Ikem A. Measurement of volatile organic compounds in bottled and tap waters by purge and trap GC–MS: are drinking water types different? J Food Compos Anal. 2010;23(1):70–7.

    Article  CAS  Google Scholar 

  • ISO. Water quality-Sampling- Part 3: preservation and handling of water samples. Geneva: International Organization for Standardization (ISO) 5667-3:2012(E); 2012.

    Google Scholar 

  • Jurdakova H, Kubinec R, Jurcisinova M, Krkosova Z, Blasko J, Ostrovsky I, Sojak L, Berezkin VG. Gas chromatography analysis of benzene, toluene, ethylbenzene, and xylenes using newly designed needle trap device in aqueous samples. J Chrom A. 2008;1194:161–4.

    Article  CAS  Google Scholar 

  • Kountouriotis A, Aleiferis PG, Charalambides AG. Science of the total environment numerical investigation of VOC levels in the area of petrol stations. Sci Total Environ. 2014;470-471:1205–24.

    Article  CAS  Google Scholar 

  • Kubinec R, Berezkin VG, Górová R, Addová G, Mracnová H, Soják L. Needle concentrator for gas chromatographic determination of BTEX in aqueous samples. J Chromatogr B. 2004;800:295–301.

    Article  CAS  Google Scholar 

  • Leivadara SV, Nikolaou AD, Lekkas TD. Determination of organic compounds in bottled waters. Food Chem. 2008;108(1):277–86.

    Article  CAS  Google Scholar 

  • Mirzaei A, Leonardi SG, Neri G. Detection of hazardous volatile organic compounds (VOCs) by metal oxide nanostructures-based gas sensors. A Rev Cer Inter. 2016;42:15119–41.

    Article  CAS  Google Scholar 

  • RCWHC. (2005) Regulation concerning water intended for human consumption. Official Gazette: 17.02.2005. Number: 25730.

    Google Scholar 

  • Rozet E, Marini RD, Ziemons E, Boulanger B, Hubert P. Advances in validation, risk and uncertainty assessment of bioanalytical methods. J Pharm Biomed Anal. 2011;55(4):848–58.

    Article  CAS  Google Scholar 

  • Tehrani R, Van Aken B. Hydroxylated polychlorinated biphenyls in the environment: sources, fate, and toxicities. Environ Sci Pollut Res. 2014;21(10):6334–45.

    Article  CAS  Google Scholar 

  • Tombs MC. Volatile organic compounds in water: gas chromatography, Encyclopedia of Separation Science. Warrington: North West Water Limited; 2000.

    Google Scholar 

  • Tsuchiya Y (2010) Organic chemicals as contaminants of water bodies and drinking water, Water Quality and Standards. Volume II.

  • Ueta I, Abd Razak N, Mizuguchi A, Kawakubo S, Saito Y, Jinno K. Needle-type extraction device for the purge and trap analysis of 23 volatile organic compounds in tap water. J Chrom A. 2013;1317:211–6.

    Article  CAS  Google Scholar 

  • WHO. Indoor air quality: organic pollutants. Report on a WHO Meeting. Berlin: World Health Organization; 1987.

    Google Scholar 

  • WHO. Guidelines for drinking water quality. Drinking water standards. A Public Health Perspective. Geneva: World Health Organization; 1993.

    Google Scholar 

  • WHO. Guidelines for drinking-water quality. A Public Health Perspective. Geneva: World Health Organization; 2001.

    Google Scholar 

  • Wu Z, Fung YS. Isolation and determination of volatile organic compounds from water by dynamic purge-and-trap technique coupled with capillary gas chromatography. Int J Environ Anal Chem. 2010;82:431–42.

    Article  Google Scholar 

Download references

Acknowledgements

This study was supported within the scope of Hazardous Waste Analysis Project situated in TUBITAK MAM Environment and Cleaner Production Institute. The authors wish to thank H. Demir and E. Ay for their supports in the laboratory.

Funding

Not applicable.

Author information

Authors and Affiliations

Authors

Contributions

OC conceived of the study and carried out the design of experiment. BG carried out the sample preparation and analysis, assessment of data, and helped to draft the manuscript. The author(s) read and approved the final manuscript.

Ethics approval and consent to participate

Not applicable.

Corresponding author

Correspondence to Barış Güzel.

Ethics declarations

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Güzel, B., Canli, O. Method validation and measurement uncertainty of possible thirty volatile organic compounds (VOCs) presented in the polyethylene present in bottled drinking waters sold in Turkey. J Anal Sci Technol 11, 44 (2020). https://doi.org/10.1186/s40543-020-00242-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40543-020-00242-6

Keywords