Skip to main content

Table 1 Brain areas showing significantly increased quantitative anisotropy (QA) values with high-dielectric pad compared to those without it

From: Benefits of high-dielectric pad for neuroimaging study in 7-Tesla MRI

Regions

Pad

p-Value

No. fiber tracks

Without

With

Whole brain

0.36 ± 0.19

0.56 ± 0.26

0.003**

6690

Corticospinal tract (CST)

0.15 ± 0.07

0.27 ± 0.21

0.150

2997

Frontopontine tract (FPT)

0.23 ± 0.16

0.40 ± 0.25

0.184

1863

Splenium of the corpus callosum (SCC)

0.49 ± 0.26

0.76 ± 0.32

0.088

815

Fornix (FX)

0.37 ± 0.20

0.64 ± 0.20

0.035*

517

Inferior fronto-occipital fasciculus (IFOF)

0.35 ± 0.18

0.55 ± 0.25

0.044*

240

Cerebellum (CB)

0.14 ± 0.06

0.22 ± 0.13

0.111

121

Middle cerebellar peduncle (MCP)

0.19 ± 0.11

0.33 ± 0.24

0.170

105

Body of the corpus callosum (BCC)

0.52 ± 0.25

0.56 ± 0.25

0.656

32

  1. p-Values of Paired-samples t test
  2. *Statistically significant with p-value < 0.05
  3. **Statistically significant with p-value < 0.01