Skip to main content

Table 2 Intermediate precision by method validation

From: Enhancement of teprenone isomer separation by subcritical fluid chromatography using porous graphitic carbon column

Validation

Concentration (μg/mL)

Measured conc. (μg/mL) (95% CI*)

Precision (RSD, %)

Accuracy (%) (95% CI*)

Intra-day (n = 6)

500.0

494.8 ± 1.5

0.4

99.0 ± 0.3

Inter-day (n = 3)

Day 1

500.0

495.7 ± 3.4

0.6

99.1 ± 0.6

Day 2

500.0

496.5 ± 1.8

0.3

99.2 ± 0.3

Day 3

500.0

494.7 ± 7.7

1.4

98.9 ± 1.5

  1. *CI confidence interval