Skip to main content

Table 2 The recoveries of serum samples spiked with HER-2 and HE4 by dual-label TRFIA

From: Dual-label time-resolved fluoroimmunoassay for simultaneous measurement of human epidermal growth factor receptor 2 and human epididymis protein 4 in serum

Sample

HER-2

HE4

Spiked/ng ml−1

Observed/ng ml−1, n = 3

Recovery (%)

CV (%)

Spiked/pmol l−1

Observed/pmol l−1, n = 3

Recovery (%)

CV (%)

Case 1

0

9.8

  

0

48.1

  

20

27.6

89

10.1

50

102.3

108.4

5.2

50

57.5

95.4

3.4

100

142.1

94

7.1

100

106.1

96.3

2.6

200

229.7

90.8

3.6

Case 2

0

16.0

  

0

31.0

  

20

34.9

94.5

7.3

50

80.2

98.4

2.9

50

66.7

101.4

1.5

100

123.9

92.9

4.9

100

109.6

93.6

5.2

200

237.6

103.3

2.8

Case 3

0

17.5

  

0

37.0

  

20

35.6

90.5

2.6

50

87.8

101.6

9.5

50

63.4

91.8

3.0

100

127.3

90.3

2.6

100

115.2

97.7

9.1

200

215.8

89.4

3.4

Average

  

94.5

5.0

  

96.6

4.7