Table 8 Results of effect of variation in column temperature

From: A novel UPLC-PDA isocratic method for the quantification fulvestrant in oil-based pre-filled syringe injection matrix formulations

S. No. Name of the parameter 25 °C 30 °C
1 Tailing factor for fulvestrant peak 0.93 0.94
2 % RSD of fulvestrant peak areas for five replicate injections of standard 0.1 0.1
3 USP plate count for fulvestrant peak 4013 3980
4 % Assay 99.3 99.8