From: A novel UPLC-PDA isocratic method for the quantification fulvestrant in oil-based pre-filled syringe injection matrix formulations
S. No.
Name of the parameter
25 °C
30 °C
1
Tailing factor for fulvestrant peak
0.93
0.94
2
% RSD of fulvestrant peak areas for five replicate injections of standard
0.1
3
USP plate count for fulvestrant peak
4013
3980
4
% Assay
99.3
99.8