Skip to main content

Table 10 Results of effect of variation in organic (acetonitrile) variation

From: A novel UPLC-PDA isocratic method for the quantification fulvestrant in oil-based pre-filled syringe injection matrix formulations

S. No.

Name of the parameter

Water:ACN:MeOH (300:400:300)

Water:ACN:MeOH (300:360:300)

Water:ACN:MeOH (300:440:300)

1

Tailing factor for fulvestrant peak

0.94

0.94

0.95

2

% RSD of fulvestrant peak areas for five replicate injections of standard

0.2

0.1

0.3

3

USP plate count for fulvestrant peak

4229

4263

4002

4

% Assay

101.4

100.7