Skip to main content

Advertisement

Table 8 System suitability parameters

From: Development of a new, sensitive, and robust analytical and bio-analytical RP-HPLC method for in-vitro and in-vivo quantification of naringenin in polymeric nanocarriers

Parameters Obtained values Standard values
Theoretical plate 9189 ≥ 2000
Asymmetry 1.1 < 2 (0.9–1.1)
LOD (μg/ml) 0.1 μg/ml 3* (signal/noise)
LOQ (μg/ml) 0.44 μg/ml 10 * (signal/noise)
Retention time 7.1 > 2
Precision 0.5–1.2 (intra-day and inter-day) RSD ≤ 2
Accuracy 0.28–1.19 RSD ≤ 2