Skip to main content

Table 8 System suitability parameters

From: Development of a new, sensitive, and robust analytical and bio-analytical RP-HPLC method for in-vitro and in-vivo quantification of naringenin in polymeric nanocarriers

Parameters

Obtained values

Standard values

Theoretical plate

9189

≥ 2000

Asymmetry

1.1

< 2 (0.9–1.1)

LOD (μg/ml)

0.1 μg/ml

3* (signal/noise)

LOQ (μg/ml)

0.44 μg/ml

10 * (signal/noise)

Retention time

7.1

> 2

Precision

0.5–1.2 (intra-day and inter-day)

RSD ≤ 2

Accuracy

0.28–1.19

RSD ≤ 2