Table 1 Linear regression and selectivity data of Pb, Cd, and Cr for aqueous and matrix matching calibration curves

From: Development and validation of a GF-AAS method and its application for the trace level determination of Pb, Cd, and Cr in fish feed samples commonly used in the hatcheries of Bangladesh

Analyte

Linear range (μg/L)

Calibration with aqueous standard solutions

Calibration with matrix matched standard solutions

Slope ± SD (μg/L)−1

Intercept ± SD

Correlation Coefficient

Slope ± SD (μg/kg)−1

Intercept ± SD

Correlation coefficient

Pb

15–75

0.00084 ± 0.00002

0.00053 ± 0.00025

0.998

0.00086 ± 0.00002

0.00521 ± 0.00164

0.998

Cd

1–6

0.01941 ± 0.00019

0.01430 ± 0.00374

0.997

0.01984 ± 0.00065

0.03423 ± 0.00319

0.997

Cr

4–20

0.00607 ± 0.00002

0.02070 ± 0.00479

0.998

0.00615 ± 0.00003

0.00463 ± 0.002811

0.998