Skip to main content

Table 8 Robustness of the method by changing wavelength ( n= 3) (concentration 100μg/mL)

From: Stability-indicating RP-HPLC method for the determination of 6-gingerol in polyherbal formulations

Detection wavelength (nm)

Mean area ± SD

Mean R t ± SD

%RSD of area

%RSD of R t

Actual

Used

Level

280

278

−2

9,823,452 ± 77,546

13.1 ± 0.26

0.78

1.9

280

0

9,921,953 ± 65,391

13.3 ± 0.20

0.89

0.66

282

+2

9,895,383 ± 100,166

13.6 ± 0.23

1.01

1.7