Skip to main content

Table 1 Intra- and inter-day accuracy, precision and reinjection reproducibility of the method

From: Development and validation of a sensitive assay for the quantification of arachidonoylcyclopropylamide (ACPA) in cell culture by LC–MS/MS

Concentration (ng/mL)

Intra-day (n = 3)

Inter-day (n = 3)

Reinjection reproducibility (n = 6)

Accuracy

(Bias, %)

Precision

(CVa, %)

Accuracy

(Bias, %)

Precision

(CVa, %)

Accuracy

(Bias, %)

Precision

(CVa, %)

1.80

1.48

1.07

 − 1.25

3.12

1.73

0.92

31.25

 − 0.23

1.97

 − 0.17

1.07

1.67

1.79

250.00

1.49

1.73

 − 0.37

0.09

1.66

1.37

500.00

 − 0.23

0.77

0.42

0.79

3.36

0.14

  1. aCV Coefficient of variation