Test parameter | Typical acceptance criteria | 2-Bromoethanol (2-BE) | 2-Bromoethylmethanesulfonate (2-BEM) |
---|---|---|---|
System suitability | RSD (%) for peak area response (n = 6) | 1.27% | 0.28% |
Cumulative RSD (%) for peak area response (all injections) | 1.10% | 3.15% | |
System suitability (for Intermediate precision) | RSD (%) for peak area response (n = 6) | 0.36% | 1.44% |
Cumulative RSD (%) for peak area response (all injections) | 0.83% | 1.43% | |
Specificity | Blank interference and USP Resolution | No blank interference, 2-BE and 2-BEM are well resolved from residual solvent and matrix peaks | |
Sensitivity | Concentration | LOD—8.1 ppm LOQ—24.3 ppm | LOD—8.0 ppm LOQ—24.1 ppm |
S/N for LOD solution should be > 3:1 | 4:1 | 3:1 | |
S/N for LOQ solution should be > 10:1 | 13:1 | 11:1 | |
RSD (%) for six replicate injections of LOQ solution should be ≤ 15.0% | 2.61% | 1.09% | |
Linearity | Range | 24.1–3013.5 ppm | 25.1–301.4 ppm |
Calibration Equation | y = 0.7513 x − 0.0538 | y = 0.8364x − 0.1050 | |
r2 | 0.9999 | 0.9999 | |
Residual plots | Random scatter | Random scatter | |
Accuracy | Average recovery (n = 3) from the spiked samples performed at 5 levels should be between 80 and 100%; RSD (%) should be ≤ 10.0% | LOQ—95.8%; 4.75% 50–95.9%; 2.16% 100–99.8%; 3.60% 150–94.5%; 1.01% 300–101.1%; 0.29% | LOQ—93.5%; 1.27% 50–90.6%; 4.50% 100–104.4%; 0.67% 150–95.7%; 0.40% 300–96.7%; 0.43% |
Precision | RSD (%) for six preparations at 100% spike level should be ≤ 10.0% | 1.39% | 2.73% |
Intermediate Precision | RSD (%) for six preparations at 100% spike level should be ≤ 10.0% | 1.33% | 2.35% |
Robustness 0.9 mL min−1 Flow | RSD (%) for peak area response (n = 6) %Recovery (n = 3) for 100% spiked solution | 0.39% 97.6% | 1.70% 93.5% |
Robustness 1.1 mL min−1 Flow | RSD (%) for peak area response (n = 6) %Recovery (n = 3) for 100% spiked solution | 1.04% 91.3% | 0.95% 96.9% |
Robustness 45° Oven Temp | RSD (%) for peak area response (n = 6) %Recovery (n = 3) for 100% spiked solution | 0.47% 93.7% | 1.35% 94.9% |
Robustness 55° Oven Temp | RSD (%) for peak area response (n = 6) %Recovery (n = 3) for 100% spiked solution | 1.02% 95.9% | 0.71% 95.3% |