Skip to main content

Table 1 Intra- and inter-day precision and accuracy, recovery and matrix effect of mefatinib in human plasma (n = 6)

From: Development and validation of a new LC–MS/MS method for the determination of mefatinib in human plasma and its first application in pharmacokinetic studies

Concentration (ng mL−1)

Intra-day

Inter-day

Recovery

Matrix effect

Mean ± SD (ng mL−1)

Accuracy (Bias %)

Precision (CV %)

Mean ± SD (ng mL−1)

Accuracy (Bias %)

Precision (CV %)

Mean ± SD

Precision (CV %)

Mean ± SD

Precision (CV %)

0.500 (LLOQ)

0.470 ± 0.0129

− 6.00

2.74

0.485 ± 0.0202

− 3.00

4.16

NA

NA

NA

NA

1.00 (LQC)

0.976 ± 0.0363

− 2.40

3.72

0.971 ± 0.0312

− 2.90

3.21

112.09 ± 2.20

1.96

1.02 ± 0.02

1.83

16.0 (MQC)

15.3 ± 0.297

− 4.38

1.94

15.2 ± 0.248

− 5.00

1.63

112.48 ± 2.50

2.22

NA

NA

160.0 (HQC)

157 ± 3.02

− 2.13

1.93

155 ± 2.68

− 3.13

1.73

113.19 ± 3.17

2.80

1.01 ± 0.01

1.33