Test parameter | Typical acceptance criteria | Result | ||
---|---|---|---|---|
PGI-1 | PGI-2 | PGI-3 | ||
System suitability | RSD (%) for peak area response (n = 6) | 0.97% | 1.62% | 1.94% |
Specificity | Interference from blank | No interference | No interference | No interference |
Sensitivity | Concentration | LOD–2.5 ppm LOQ–8.0 ppm | LOD–3.5 ppm LOQ–11.0 ppm | LOD–4.0 ppm LOQ–12.5 ppm |
RSD for six replicate injections of LOQ solution should be ≤ 15.0% | 4.13% | 4.21% | 3.29% | |
Linearity | Range | 2.5–100 ppm | 3.5–100 ppm | 4.0–100 ppm |
Calibration equation | y = 130.97x + 178.5 | y = 123.16x − 41.816 | y = 69.578x−36.595 | |
R2 | 0.9992 | 0.9969 | 0.9997 | |
Accuracy | Average recovery (n = 3) from the spiked samples performed at 3 levels—LOQ–150%; RSD should be ≤ 10.0% | LOQ–96.9%; 1.28% 100%–100.1%; 1.88% 150%–100.5%; 0.97% | LOQ–98.9%; 1.98% 100%–99.8%; 2.08% 150%–99.0%; 1.11% | LOQ–99.3%; 3.57% 100%–98.8%; 1.36% 150%–97.7%; 1.35% |
Precision (analyst 1) | RSD (%) for six preparations at 100% level should be ≤ 10.0% | 2.37% | 3.15% | 2.94% |
Intermediate precision (analyst 2) | RSD (%) for six preparations at 100% level should be ≤ 10.0% | 2.96% | 3.42% | 3.83% |
Solution stability | Standard and 100% spiked solution stored at ambient laboratory conditions (25 ± 5 °C) and refrigerated conditions (2–8 °C) were studied for 48 h | Stable for 48 h | Stable for 48 h | Stable for 48 h |
Robustness | RSD (%) for peak area response (n = 6) with 0.7 mL/min flow rate %Recovery (n = 3) for 100% spiked solution | 1.28% 100.1% | 1.94% 98.7% | 1.82% 96.8 |
RSD (%) for peak area response (n = 6) with 1.0 mL/min flow rate %Recovery (n = 3) for 100% spiked solution | 4.1% 96.5% | 3.4% 96.8% | 3.5% 95.2% | |
RSD (%) for peak area response (n = 6) with 12 V cone voltage %Recovery (n = 3) for 100% spiked solution | 1.34% 99.6% | 1.9% 97.7% | 2.2% 97.4% | |
RSD (%) for peak area response (n = 6) with 18 V cone voltage %Recovery (n = 3) for 100% spiked solution | 3.1% 95.3% | 4.3% 94.7% | 4.9% 93.8% |