Table 1 Analytical performances of the proposed method at model sample solution and plasma matrix

From: A new liquid phase microextraction method-based reverse micelle for analysis of dexketoprofen in human plasma by HPLC-DAD

Model sample solution Plasma sample
Parameter Value Parameter Value
LDA, μg/mL 0.01-10.00 LDA, μg/mL 0.05–2.00
Intraday precision, % RSD (n = 6) 1.2 Intraday precision, % RSD (n = 6) 1.7
Interday precision, % RSD (n = 6) 2.6 Interday precision, % RSD (n = 6) 3.9
r2 0.9999 r2 0.9996
LOD, ng/mL 0.8 LOD, ng/mL 12.8
LOQ, ng/mL 2.4 LOQ, ng/mL 38.8
EF 20 EF 2