Skip to main content

Table 7 Summary of linear regression data and validation parameters of HPTLC method

From: Application of Plackett-Burman and central composite designs for screening and optimization of factor influencing the chromatographic conditions of HPTLC method for quantification of efonidipine hydrochloride

Parameters

HPTLC

Linearity

 Range (ng/band)

300–1800

 Slope

4.812

 Intercept

506.5

 Correlation coefficient

0.999

Accuracy

 % recovery (% RSD)

98.57–101.40%, 0.34–0.75

Precision

 Intra-day precision (n = 3) (% RSD)

0.41–1.54

 Inter-day precision (n = 3) (% RSD)

1.14–1.66

 Repeatability (n = 6) (% RSD)

0.65

Sensitivity

 

 Limit of detection (ng)

10.41

 Limit of quantification (ng)

31.57

Robustness

Robust

Ruggedness

 Analyst-I (mean ± SD)

101.63 ± 0.45

 Analyst-I (mean ± SD)

101.69 ± 0.70

Analysis of pharmaceutical preparation (mean ± SD)

100.03 ± 1.63

  1. n Number of determinations