Skip to main content

Table 7 The measurement result of Au element in soil samples of actual Au mining area

From: Gold determination in soil by ICP-MS: comparison of sample pretreatment methods

Sample

Method

Average value (ng/g)

RSD (%)

added amount (ng/g)

Measurement value (ng/g)

Recovery rate (%)

A1

I

180.47

1.8

200.00

373.02

96.3

II

190.19

3.4

200.00

399.04

104.4

A2

I

97.26

1.7

100.00

195.59

98.3

II

103.64

2.1

100.00

200.97

97.3

A3

I

74.21

1.3

100.00

175.88

101.7

II

78.09

0.7

100.00

177.39

99.3

B1

I

4.89

2.9

5.00

9.73

96.8

II

5.20

2.3

5.00

10.08

97.6

B2

I

2.45

1.2

2.00

4.36

95.5

II

2.56

3.6

2.00

4.50

96.5

B3

I

1.09

5.1

1.00

2.10

101.0

II

1.13

3.1

1.00

2.12

99.0

  1. I electric heating plate digestion method, II microwave digestion method, RSD relative standard deviation