Skip to main content

Table 6 The measurement result of Au element in five CRMs with two digestion methods

From: Gold determination in soil by ICP-MS: comparison of sample pretreatment methods

CRM

Standard value (ng/g)

Method

Average value (ng/g)

ΔlgC (%)

RE (%)

RSD (%)

GBW07805

0.85

I

0.82

1.6

3.5

6.2

II

0.87

1.0

2.4

1.8

GBW07806

3.10

I

2.97

1.9

4.2

3.4

II

3.29

2.6

6.1

1.7

GBW07246

20.80

I

21.72

1.9

4.4

5.1

II

23.03

4.4

10.7

3.1

GBW07247

50.00

I

50.18

0.2

0.4

2.3

II

53.89

3.3

7.8

1.3

GBW07248

100.00

I

105.50

2.3

5.5

0.7

II

110.35

4.3

10.4

1.7

  1. I electric heating plate digestion method, II microwave digestion method, ΔlgC logarithmic deviation, RE relative error, RSD relative standard deviation