Skip to main content

Table 3 Summary of stability results

From: Accurate quantification of EVT201 and its two metabolites in human urine using UHPLC-MS/MS method with solid phase extraction

Compound

Parameter

Benchtop stability (24 h)

Autosampler stability (192 h)

Freeze-thaw stability

Long-term stability

LQC (0.6 ng/mL)

HQC (160 ng/mL)

LQC (0.6 ng/mL)

HQC (160 ng/mL)

Cycles

LQC (0.6 ng/mL)

HQC (160 ng/mL)

Days

LQC (0.6 ng/mL)

HQC (160 ng/mL)

EVT201

Mean

0.618

146

0.607

143

10

0.673

181

944

0.595

150

RE%

3.0

− 8.8

1.2

− 10.6

12.2

13.1

− 0.8

− 6.3

SD

0.005

3

0.011

4

0.068

8

0.057

2

CV%

0.8

2.1

1.8

2.8

10.1

4.4

9.6

1.3

Ro46-1927

Mean

0.561

137

0.661

170

9

0.596

172

939

0.673

181

RE%

− 6.5

− 14.4

10.2

6.3

− 0.7

7.5

12.2

13.1

SD

0.079

14

0.019

5

0.008

13

0.068

8

CV%

14.1

10.2

2.9

2.9

1.3

7.6

10.1

4.4

Ro18-5528

Mean

0.557

139

0.622

144

10

0.529

153

694

0.593

172

RE%

− 7.2

− 13.1

3.7

− 10.0

− 11.8

− 4.4

− 0.7

7.5

SD

0.021

5

0.036

2

0.020

2

0.008

13

CV%

3.8

3.6

5.8

1.4

3.8

1.3

1.3

7.6