Table 3 Extraction recoveries of lenalidomide and IS in human plasma

From: Development and validation of a liquid chromatography-tandem mass spectrometric method for the determination of lenalidomide in human plasma and its application on bioequivalence studies

Spiked conc. (ng/mL)Extraction recoveries (%)% CV
26.2597.322.15
323.0698.580.59
807.6498.282.12
Mean98.061.62