Skip to main content

Table 3 Extraction recoveries of lenalidomide and IS in human plasma

From: Development and validation of a liquid chromatography-tandem mass spectrometric method for the determination of lenalidomide in human plasma and its application on bioequivalence studies

Spiked conc. (ng/mL)

Extraction recoveries (%)

% CV

26.25

97.32

2.15

323.06

98.58

0.59

807.64

98.28

2.12

Mean

98.06

1.62

Internal standard

88.78

4.00