Skip to main content

Table 1 Precision and accuracy results for lenalidomide in human plasma

From: Development and validation of a liquid chromatography-tandem mass spectrometric method for the determination of lenalidomide in human plasma and its application on bioequivalence studies

Spiked conc. (ng/mL)

Intra-day

Inter-day

Calculated conc. (mean ± SD)

Precision (% CV)

Accuracy (% RE)

Calculated conc. (mean ± SD)

Precision (% CV)

Accuracy (% RE)

10.00

10.76 ± 0.38

3.51

− 7.58

10.60 ± 0.31

2.88

− 6.01

26.25

28.09 ± 0.41

1.47

− 7.0

28.08 ± 0.83

2.95

− 6.96

323.06

326.65 ± 14.93

4.57

− 1.11

324.68 ± 12.76

3.93

− 0.50

807.64

791.12 ± 20.57

2.60

2.05

802.87 ± 33.89

4.22

0.59