Skip to main content

Table 4 Results of intra- and inter-day precision and accuracy

From: Development of membrane electrodes for selective determination of lisinopril in pharmaceuticals

ISE

LNP taken, mmol l−1

Intra-day variations

Inter-day variations

LNP found*, mmol l−1

%RSD

%RE

LNP found$, mmol l−1

%RSD

%RE

LNP-PTA

0.075

0.073

3.81

2.67

0.077

3.89

2.67

0.500

0.509

4.89

1.80

0.511

4.11

2.20

1.000

1.01

2.97

1.00

1.020

4.33

2.00

LNP-PMA

0.075

0.072

2.91

4.00

0.073

2.99

2.67

0.500

0.494

4.19

1.20

0.516

3.15

3.20

1.000

1.020

3.97

2.32

1.030

3.63

3.00

  1. *Mean value of seven measurements; $Mean value of five measurements