Table 7 Results of effect of variation in flow rate

From: A novel UPLC-PDA isocratic method for the quantification fulvestrant in oil-based pre-filled syringe injection matrix formulations

S. No.

Name of the parameter

0.3 mL min−1

0.24 mL min−1

0.36 mL min−1

1

Tailing factor for fulvestrant peak

0.94

0.93

0.97

2

% RSD of fulvestrant peak areas for five replicate injections of standard

0.1

0.1

0.1

3

USP plate count for fulvestrant peak

4013

4054

3970

4

% Assay

99.3

101.4

99.5