Table 5 Linearity of detector response of fulvestrant

From: A novel UPLC-PDA isocratic method for the quantification fulvestrant in oil-based pre-filled syringe injection matrix formulations

% Level

Concentration (ppm)

Area

Average area

% RSD

80%-1

79.8540

657,899

659,836

0.2

80%-2

79.8540

660,890

80%-3

79.8540

659,817

90%-1

89.8357

742,007

741,004

0.1

90%-2

89.8357

740,886

90%-3

89.8357

740,119

100%-1

99.8175

824,818

824,416

0.1

100%-2

99.8175

824,942

100%-3

99.8175

823,487

110%-1

109.7992

910,473

909,937

0.1

110%-2

109.7992

910,860

110%-3

109.7992

908,479

120%-1

119.7810

992,290

993,041

0.3

120%-2

119.7810

990,134

120%-3

119.7810

996,701

Correlation coefficient

0.9999

Intercept

− 9696.0344

Slope

8368.7029

Bias at 100%

− 1.18

Standard error

1291.356784