Table 4 Precision study

From: Design of experiment avenue for development and validation of RP-HPLC-PDA method for determination of apremilast in bulk and in in-house tablet formulation

Standard concentration (μg/ml)

Amount found (μg/ml)

% Amount found (μg/ml)

(n = 3)

% RSD

Intra-day precision

  4

3.98

99.69

0.15

  8

8.07

100.89

0.31

  10

10.14

101.44

0.77

Inter-day precision

  4

3.94

98.61

0.17

  8

8.07

100.87

0.82

  10

9.56

98.63

0.98

  1. n number of determinations