Skip to main content

Advertisement

Table 2 ANOVA results

From: Design of experiment avenue for development and validation of RP-HPLC-PDA method for determination of apremilast in bulk and in in-house tablet formulation

Responses ± SD Mean % C. V. Press r 2 Adjusted R-square Predicted R-square Adequate precision p value
Retention time 0.58 5.81 9.96 37.14 0.873 0.7103 1.0107 7.437 0.0188
Theoretical plates 29.11 2280.76 1.28 61,942.00 0.977 0.9491 0.7675 17.25 0.0001
Tailing factor 0.026 1.08 2.41 0.051 0.9917 0.9810 0.9105 29.46 0.0001