Table 3 Tolerance limits for interferences in determination of 50.0 μg L−1 Cu(II) ions

From: Determination of copper(II) by flame atomic absorption spectrometry after its perconcentration by a highly selective and environmentally friendly dispersive liquid–liquid microextraction technique

Foreign species

Tolerated ratio (Wspecies/WCu(II))

NO3, Cl, BrO3, ClO4, S2O82−, Cr3+, Cr6+, Mn2+, Li+, Co2+, K+, Al3+, Pd2+, Ni2+, Ag+

5000

PO43−, IO4, F, SO32−, citrate, S2O52−, SCN, CN, HCOO, acetate, CO32−, C2O42−, Ba2+, Zn2+, Fe2+, Fe3+

1000

ClO3, NO2, Cd2+

500

Hg2+

100

Sn2+, Pb2+

10