Skip to main content

Table 3 Validation results obtained from spectrofluorimetry for the quantitative determination of SIL

From: An effective and sensitive stability-indicating chromatographic approach based on HPLC for silodosin assay

Parameters

Spectrofluorimetry

Wavelength (nm)

λex = 330, λem = 460

Linearity rangea (μM)

0.2–36.4

Slopeb

26.993

Interceptb

4.4526

Correlation coefficient (R 2)

0.9975

LOD (nM)

7.70

LOQ (nM)

26.80

Repeatability of peak intensity (RSD %)c

0.48

Repeatability of peak intensity (RSD %)d

0.85

  1. aAverage of three determinations
  2. b I FL (a.u.) = m C SL (μM) + b where C is the concentration and I FL is the fluorescence intensities
  3. cIntra-day measurements—each value is the mean of five experiments
  4. dInter-day measurements—each value is the mean of five experiments