Skip to main content

Table 7 Extended accuracy and precision analysis

From: A LC–MS/MS method with column coupling technique for simultaneous estimation of lamivudine, zidovudine, and nevirapine in human plasma

Analyte

QC sample

Actual conc. (ng/ml)

Estimated conc. (ng/ml)

Accuracy (%)

% CV

Lamivudine

LQC

27.53

26.88

97.65

2.72

HQC

1146.90

1153.16

100.55

2.52

Zidovudine

LQC

29.84

29.07

97.42

4.71

HQC

1243.43

1211.79

97.46

2.12

Nevirapine

LQC

44.29

44.85

101.25

2.28

HQC

1845.48

1815.04

98.35

2.47