Skip to main content

Table 3 Detection and quantification limits for ifosfamide, cyclophosphamide, and paclitaxel

From: Accessible analytical methodology for assessing workplace contamination of antineoplastic drugs in limited-resource oncology health-care settings

 

Limit of detection

Limit of quantification

Antineoplastic drug

Instrumental (ng/mL)a

Wipe media (ng/cm2)b

Concentrated wipe media (ng/cm2)c

Instrumental (ng/mL)a

Wipe media (ng/cm2)b

Concentrated wipe media (ng/cm2)c

Ifosfamide

10

0.2

0.02

32

0.8

0.08

Cyclophosphamide

41

1

0.1

124

3

0.3

Paclitaxel

13

0.3

0.03

39

1

0.1

  1. aCalculated as described in “Sensitivity” section
  2. bBased on the 400-cm2 wipe area with extraction into 10 mL, no further concentration steps
  3. cBased on the 400-cm2 wipe area with extraction into 10 mL, tenfold further concentration step