Skip to main content

Table 1 Tolerant limits of coexisting ions for the extraction of 0.5 μg mL−1 silver ions

From: Ultrasound-assisted emulsification microextraction and preconcentration of trace amounts of silver ions as a cyclam complex

Interference ion

Added as

Tolerance limit ion ratio

(Cion/CAg)

% Recovery

Mg2+

Mg(NO3)2

1000

96 ± 3

K+

KNO3

100

95.1 ± 2.2

Ca2+

Ca(NO3)2.4H2O

10

95.4 ± 2.5

Co2+

Co(NO3)2.6H2O

10

97.1 ± 3.4

Cd2+

Cd(NO3)2.6H2O

10

96.0 ± 1.5

Hg+

Hg2(NO3)2

5

95.4 ± 2.3

Al3+

Al(NO3)3

5

97.32 ± 2.4

Na+

NaNO3

5

96.3 ± 3.3

Mn2+

Mn(NO3)2.4H2O

5a

95.0 ± 2.3

Pb2+

Pb(NO3)2

1

94.0 ± 3.5

Hg2+

Hg(NO3)2

1a

98.8 ± 1.4

Ni2+

Ni(NO3)2

5a

95.35 ± 2.5

Fe3+

Fe(NO3)3.9H2O

1b

95.76 ± 3.4

NO3

KNO3

100

95.2 ± 3.5

CO3 2−

K2CO3

100

95.0 ± 4.3

OAC

CH3COOK

25

100.4 ± 1.3

Cl

KCl

10

95.06 ± 3.2

  1. aThe interference effect in the presence of EDTA