Skip to main content

Table 2 Accuracy and precision data for the determination of KT by the proposed methods

From: Spectrophotometric determination and spectroscopic studies on Schiff base and charge transfer complex of ketorolac tromethamine

Methods

KETO taken (μg mL−1)

Intraday (n = 5)

Interday (n = 5)

 

KETO founda ± SD (μg mL−1)

RE (%)

RSD (%)

KETO founda ± SD (μg mL−1)

RE (%)

RSD (%)

Method A

2.5

2.47 ± 0.02

1.20

0.81

2.51 ± 0.03

0.40

1.52

 

6.5

6.46 ± 0.02

0.61

0.30

6.51 ± 0.06

0.27

1.04

 

10.5

10.51 ± 0.05

0.09

0.48

10.47 ± 0.06

0.21

0.61

Method B

10

10.04 ± 0.08

0.40

0.79

10.01 ± 0.13

0.10

1.29

 

26

26.02 ± 0.12

0.07

0.46

25.88 ± 0.11

0.46

0.42

 

34

34.02 ± 0.12

0.05

0.35

33.82 ± 0.08

0.52

0.23

  1. SD standard deviation, RSD relative standard deviation, RE relative error
  2. aMean value of five determinations