From: Development and validation of RP-HPLC method for glimepiride and its application for a novel self-nanoemulsifying powder (SNEP) formulation analysis and dissolution study
Actual conc. (μg/mL)
Calculated conc. (μg/mL) ± SD (n = 3)
%RSD
%Recovery
1.2
1.1983 ± 0.00153
0.127
99.87
1.4
1.403 ± 0.01058
0.754
100.3