Skip to main content

Table 3 Evaluation of accuracy of the investigated analytical procedure at three concentration levels within the specified range

From: Spectrofluorimetric determination of certain antidepressant drugs in human plasma

Drug 50.0 ng mL-1 Recovery %a 250.0 ng mL-1 450.0 ng mL-1
Fluoxetine 100.94 ± 1.29 99.80 ± 0.51 100.41 ± 0.28
Sertraline 99.84 ± 1.69 98.91 ± 0.55 100.14 ± 0.30
  50.0 ng mL -1 Recovery % a 250.0 ngmL -1 550.0 ngmL -1
Paroxetine 100.92 ± 1.28 100.09 ± 0.67 100.06 ± 0.31
  100.0 ngmL -1 Recovery % a 350.0 ngmL -1 650.0 ngmL -1
Thioridazine 100.44 ± 1.57 101.31 ± 0.95 100.48 ± 0.41
  100.0 ngmL -1 Recovery % a 450.0 ngmL -1 750.0 ngmL -1
Amineptine 100.14 ± 1.87 98.81 ± 0.72 100.47 ± 0.44
  1. a Mean of Six replicate measurements.