Skip to main content

Table 4 Repeatability and intermediate precision of HPLC method

From: Development and validation of RP-HPLC-PDA method for the quantification of eugenol in developed nanoemulsion gel and nanoparticles

Concentration (?g?mL ?1)

Repeatability (n?=?3)

Intermediate precision (n?=?3)

Mean peak area???SD

%RSD

Mean peak area???SD

%RSD

20

468283.0???3410.6

0.73

467144.7???3398.0

0.73

100

2213959.7???11480.9

0.52

2213539.0???8008.9

0.36

600

12800060.3???45178.8

0.35

12778185.3???40768.0

0.32