Table 3 Inter- and Intra-day accuracy and precision, recovery (RE), matrix effect (ME), process efficiency (PE), and autosampler stability (AS) of target compounds in hair

From: Simultaneous determination of ethyl glucuronide, cocaine, cocaethylene, and benzoylecgonine in hair by using LC–MS/MS

Analyte

QC sample (pg/mg)

Inter-day (n = 24)

Intra-day (n = 6)

RE (n = 6) (%)

ME (n = 6) (%)

PE (n = 6) (%)

AS stability (n = 6) (%)

% bias

% CV

% bias

% CV

EtG

7

 − 9.3

14.4

2.3

10.3

 

20

0.7

4.0

4.5

4.8

89.2

91.4

81.5

109.7

 

100

 − 1.8

1.4

 − 2.7

3.7

98.5

88.0

86.7

 

500

 − 3.9

1.0

 − 5.4

5.9

104.5

96.6

100.9

101.2

COC

2

 − 3.1

3.9

1.9

5.9

 

5

3.0

0.6

4.0

1.4

100.5

81.7

82.0

96.6

 

25

3.5

1.4

2.0

0.5

100.0

81.6

81.6

 

125

 − 4.1

1.5

 − 6.1

1.6

99.6

83.1

82.8

100.8

CE

10

3.4

4.3

9.7

10.0

 

25

3.6

2.1

5.7

7.2

104.8

96.2

100.8

99.9

 

125

3.6

2.2

2.7

1.8

97.2

89.9

87.3

 

625

 − 1.5

2.6

 − 4.1

0.9

100.1

82.9

83.0

100.3

BZE

1

 − 4.4

7.7

6.4

8.4

 

2.5

3.7

3.2

8.5

1.9

97.5

105.1

102.5

99.6

 

12.5

8.3

4.8

5.1

3.9

96.6

101.9

98.4

 

62.5

 − 0.1

3.3

 − 3.4

2.8

101.7

105.4

107.1

99.9