Table 1 Intra- and inter-day accuracy and precision of the LC–MS/MS method

From: Liquid chromatography-tandem mass spectrometry for pharmacokinetics evaluation of AZD5305, a selective PARP1 inhibitor, in mice

Nominal concentration (ng/mL)

Measured concentration (ng/mL)

Accuracy (%RE)

Precision (%CV)

Mouse plasma

   

Intra-day (n = 5)

   

 1

1.01 ± 0.0615

0.917

6.09

 3

2.91 ± 0.178

3.06

6.12

 30

27.3 ± 0.917

8.94

3.36

 900

828 ± 59.9

8.00

7.24

Inter-day (n = 15)

   

 1

1.00 ± 0.101

0.172

10.1

 3

2.90 ± 0.240

3.24

8.26

 30

29.2 ± 2.03

2.78

6.94

 900

904 ± 85.6

0.444

9.47