Skip to main content

Table 4 Reproducibility and precision data evaluated through intra-day and inter-day studies

From: Development and validation of RP-HPLC method for glimepiride and its application for a novel self-nanoemulsifying powder (SNEP) formulation analysis and dissolution study

Conc. (μg/mL)

Intra-day (n = 3)

Inter-day (n = 3)

Mean peak area ± SD (n = 3)

%RSD

Mean peak area ± SD (n = 3)

%RSD

1.2

95,187 ± 605

0.63

96,391 ± 426

0.44

1.4

112,849 ± 1,077

0.95

115,782 ± 1,121

0.96

2.0

165,844 ± 1,317

0.79

169,267 ± 541

0.319